Abstract:
OBJECTIVE:This study investigated the safety and efficacy of remote programming of cochlear implants. STUDY DESIGN:Single-subject design SETTING:: Four North American clinical sites PATIENTS:: Forty cochlear implant recipients aged 12 years or older INTERVENTION:: Subjects had their cochlear implants programmed at a location that was remote from their audiologist using telecommunication with and without the support of a facilitator. MAIN OUTCOME MEASURES:Consonant-Nucleus-Consonant (CNC) word scores and the Speech, Spatial, and Qualities of Hearing Scale-C (SSQ-C) were compared using the subject's in-office MAP (program) and MAPs programmed remotely with and without the assistance of a facilitator. Additional subjective preference data were gathered from subjects and audiologists via questionnaires. RESULTS:MAPs programmed via the three different models did not yield significantly different group mean CNC word scores. No device/procedure-related adverse events occurred. SSQ-C questionnaire results indicated that recipients received similar subjective benefit from familiar in-office, remote-facilitated, and remote-unassisted MAPs. CONCLUSIONS:Remote programming is an effective means of cochlear implant service delivery. The practice was approved by the FDA on November 17, 2017 supported by the results of this study.