Powering implantable telemetry devices from localized magnetic fields

Show simple item record

dc.contributor.author McCormick, John en
dc.contributor.author Hu, Aiguo en
dc.contributor.author Nielsen, Poul en
dc.contributor.author Malpas, Simon en
dc.contributor.author Budgett, David en
dc.date.accessioned 2011-11-03T02:36:10Z en
dc.date.issued 2007 en
dc.identifier.citation 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France, 22 Aug 2007 - 26 Aug 2007. Proc of 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE. 2331-2335. 2007 en
dc.identifier.isbn 978-1-4244-0788-0 en
dc.identifier.issn 1557-170X en
dc.identifier.uri http://hdl.handle.net/2292/8599 en
dc.description.abstract This paper presents a novel method of inductively powering an implantable telemetry device over a large area. The system is based around an array of individually tuned series resonant circuits distributed across a charging pad. By varying the frequency of the driving voltage, the location of the charging field is changed. Presented are a method of controlling the resonant frequency and techniques for determining the geometry of the charging pad. Results from a nine coil system operating between 97 kHz and 209 kHz are given which can deliver 100 mW to an implanted telemeter at a height of 5cm above the charging pad. en
dc.relation.ispartof 29th IEEE Engineering in Medicine and Biology Conference en
dc.rights Items in ResearchSpace are protected by copyright, with all rights reserved, unless otherwise indicated. Previously published items are made available in accordance with the copyright policy of the publisher. en
dc.rights.uri https://researchspace.auckland.ac.nz/docs/uoa-docs/rights.htm en
dc.subject sensor-based systems for monitoring communication and minimum power consumption en
dc.title Powering implantable telemetry devices from localized magnetic fields en
dc.type Conference Item en
pubs.begin-page 2331 en
dc.rights.holder Copyright: IEEE en
dc.identifier.pmid 18002459 en
pubs.author-url https://embs.papercept.net/conferences/scripts/abstract.pl?ConfID=6&Number=1197http://ieeexplore.ieee.org.ezproxy.auckland.ac.nz/search/srchabstract.jsp?arnumber=4352793&isnumber=4352185&punumber=4352184&k2dockey=4352793@ieeecnfs&query=%28%28powering+implantable+telemetry+devices+from+localized+magnetic+fields%29%3Cin%3Eti+%29&pos=0 en
pubs.end-page 2335 en
pubs.place-of-publication Lyon en
dc.rights.accessrights http://purl.org/eprint/accessRights/RestrictedAccess en
pubs.elements-id 92495 en
pubs.org-id Bioengineering Institute en
pubs.org-id ABI Associates en
pubs.org-id Engineering en
pubs.org-id Department of Electrical, Computer and Software Engineering en
pubs.org-id Engineering Science en
pubs.org-id Medical and Health Sciences en
pubs.org-id Medical Sciences en
pubs.org-id Physiology Division en
pubs.org-id Science en
pubs.org-id Science Research en
pubs.org-id Maurice Wilkins Centre (2010-2014) en
pubs.record-created-at-source-date 2010-09-01 en
pubs.dimensions-id 18002459 en


Files in this item

Find Full text

This item appears in the following Collection(s)

Show simple item record

Share

Search ResearchSpace


Browse

Statistics